Extractive infrared multigas analyzers NDIR

Fuji Electric Extractive Gas Analyzers for process control and continuous measurement of stack emissions.
Their QAL1 certification guarantees high performance and regulatory and normative compliance.
Reliable and precise, their design quality guarantees 24-hour operation in difficult environments such as smoke from industrial chimneys.
Robustness reduces maintenance costs and increases the availability of boilers and equipment.
Extractive measurement consists of taking a gas flow from the chimney or the gas pipe via a sampling probe and conveying it to the gas analyzer.
The gas is sampled, treated in order to ensure optimal conditions for gas measurement.

Categories: , ,

Kenmerken

  • Simple maintenance thanks to self-diagnostic functions
  • Good precision and excellent stability of the measurement over time
  • Large backlit display
  • Automatic zero and span calibration
  • Correction with oxygen
  • Modbus® protocol digital link
  • Large scale dynamics (1 to 20 configurable)

Detection principle

sensor

The mass flow meter determines the amount of infrared light absorbed by the measuring cell.
Mass flow meter:

  • It converts the absorption of infrared light into an electrical signal.
  • Excellent resistance to interference thanks to the low impedance sensor.
  • The absence of moving parts makes the analyzer particularly resistant to vibrations.

Measuring components and ranges

Type ZRE ZKJ ZPA
Model ZRE - 5 components analyser - NDIR single beam technology ZKJ - 5 components gas analyser Measuring components and ranges ZPA
Range Minimum range Maximum range Minimum range Maximum range Minimum range Maximum range
NO 0-200ppm  0-5000ppm 0-50 ppm 0-5000ppm 0–200 ppm 0–5000 ppm
SO2 0-200ppm 0-10 vol% 0-50 ppm 0-10 vol% 0–200 ppm 0–10 vol%
CO2 0-100ppm 0-100 vol% 0-20 ppm 0-100 vol% 0–100 ppm 0–100 vol%
CO 0-200ppm 0-100 vol% 0-50 ppm 0-100 vol% 0–200 ppm 0–100 vol%
CH4 0-500ppm 0-100 vol% 0-200 ppm 0-100 vol% 0–500 ppm 0–100 vol%
N2O _ _ 0-200 ppm  0-2000 ppm _ _
O2
(electrochemical)
0-10 vol% 0-25 vol% _ _ 0–10 vol% 0–25 vol%
O2
(paramagnetic)
0-10 vol% 0-25 vol% 0-5 vol% 0-25 vol% 0–5 vol% 0–100 vol%
O2
(external analyser)
0-5 vol% 0-25 vol% 0-5 vol% 0-25 vol% 0–5 vol% 0–25 vol%

Fuji Electric gas analysers QAL1 lowest certified ranges

 

ZRE

ZKJ (1)

ZPA

mg/m3

ppm

mg/m3

ppm

mg/m3

ppm

CO 0 – 125 0 – 100 0 – 125 0 – 100 0 – 375 0 – 300
CO2 0 – 20 % 0 – 20 %
NO 0 – 268 (2) 0 – 200 0 – 134 (2) 0 – 100 0 – 268 (2) 0 – 200
NOx 0 – 205 (3) 0 – 100
SO2 0 – 571 0 – 200 0 – 286 0 – 100 0 – 571 0 – 200
O2 (4) 0 – 25 % 0 – 25 % 0 – 25 %

VRAAG EEN OFFERTE