Aurora 1030W

The Aurora 1030 line of TOC analyzers from OI Analytical includes instrument configurations for laboratory and process monitoring, employing 100 °C persulfate wet oxidation or thermal combustion techniques.

The Aurora 1030 line incorporates several noteworthy innovations and can be equipped with parallel reactors allowing concurrent sample processing for increased analytical throughput. The unique 1030D Dual Mode TOC analyzer allows sample processing by wet oxidation or thermal combustion using the same instrument.

Kenmerken

The Aurora 1030W TOC Analyzer uses the proven heated persulfate wet oxidation technique to analyze organic contamination levels in liquid samples. The instrument can be programmed and calibrated to analyze samples containing organic carbon at levels as low as 10 ppb and as high as 30,000 ppm. In addition to its wide range, the Aurora 1030W has an extremely high daily throughput. Up to 300 samples can be analyzed in a 24-hour period (dependent on method and application).

The reaction chamber of the Aurora 1030W is thoroughly rinsed between analyses to eliminate residue from the previous sample. This ensures the Aurora 1030W maintains the low system background necessary for high sensitivity TOC measurements. This contrasts with combustion TOC analyzers in which residual salts accumulate on catalyst surfaces, degrading oxidation efficiency and causing higher blanks and background contamination.

Tested & Approved

The heated persulfate oxidation technique has been approved and adopted in numerous regulatory compliance methods, standards, and norms for water quality testing, including: Standard Method 5310C, USEPA 415.3, USP <643> / EU 2.2.44, ASTM D 4779 and 4839, USEPA-DBPR, USEPA-SPCC, ISO 8245, and EN 1484.

Applications

The Aurora 1030W TOC Analyzer is ideal for the following applications: drinking water, pharmaceutical cleaning validation, municipal wastewater, groundwater / surface water, process water, boiler feed water and condensate, metal plating solutions, and ultrapure water.

Aurora 1030W Features:

  • Wide sample range from 10 ppb to 30,000 ppm
  • High sample efficiency – Up to 300 samples daily
  • Rinse function eliminates residue from the previous sample preventing blanks and background contamination
  • Heated persulfate oxidation technique is compliant with numerous regulatory methods, standards, and norms
  • Operates with an intuitive color touchscreen display with Windows® CE-based software
  • Options available to process more samples and test various sample types

Supported Test Methods

Method    Sample / Application
 Standard Method 5310C    Drinking Water + Wastewater
   USEPA 415.3    Drinking Water
   USP <643> / EU 2.2.44    Purified Water
   ASTM D 4779    Ultrapure Water
   ASTM D 4839    Wastewater + Seawater
   USEPA-DBPR    Disinfection Byproduct Rule
   USEPA-SPCC    Spill Prevention & Control Countermeasures
   ISO 8245    Drinking Water & Wastewater
   EN 1484    Surface & Ground Waters, Potable Water

Specificaties

Operating Principle Heated sodium persulfate oxidation
Measurement Range 10 ppb C – 30,000 ppm C (multiple calibration ranges or dilution required)
Instrument Detection LImit (IDL) 2 ppb C
Sample Injection Volume 10µL – 10mL
Method TC Acid and persulfate reaction
Method TIC Acidification with Phosphoric acid and sparging
Method TOC NPOC by heated persulfate oxidation or TC – TIC
Reagents Required Sodium persulfate, 5% phosphoric acid, rinsewater
Gas Supply N2 (99.998%), zero-grade air, or O2 (99.998%)
Measurement Technique Non-dispersive infrared (NDIR) detector
Operator Interface Color LCD touchscreen display with Windows® CE-based software
Basic Software Single instrument operation with data transfer to PC
Optional ATOC Software Network LAN/LIMS operation, data management, custom reports, 21 CFR 11 compliance
Autosampler 88 position rotary autosampler designed to fit directly underneath Aurora 1030D analyzer
Certification CE, EMC: EN61326 / Safety: IEC 61010-11 2001
Power Supply Variable voltage, 100-240VAC, 50/60Hz, 950W
Dimensions – Aurora 1030D + 1088 Autosampler 26.75 in. H x 19.5 in. W x 23 in. D.