Aurora 1030W
The Aurora 1030 line of TOC analyzers from OI Analytical includes instrument configurations for laboratory and process monitoring, employing 100 °C persulfate wet oxidation or thermal combustion techniques.
The Aurora 1030 line incorporates several noteworthy innovations and can be equipped with parallel reactors allowing concurrent sample processing for increased analytical throughput. The unique 1030D Dual Mode TOC analyzer allows sample processing by wet oxidation or thermal combustion using the same instrument.
Kenmerken
The Aurora 1030W TOC Analyzer uses the proven heated persulfate wet oxidation technique to analyze organic contamination levels in liquid samples. The instrument can be programmed and calibrated to analyze samples containing organic carbon at levels as low as 10 ppb and as high as 30,000 ppm. In addition to its wide range, the Aurora 1030W has an extremely high daily throughput. Up to 300 samples can be analyzed in a 24-hour period (dependent on method and application).
The reaction chamber of the Aurora 1030W is thoroughly rinsed between analyses to eliminate residue from the previous sample. This ensures the Aurora 1030W maintains the low system background necessary for high sensitivity TOC measurements. This contrasts with combustion TOC analyzers in which residual salts accumulate on catalyst surfaces, degrading oxidation efficiency and causing higher blanks and background contamination.
Tested & Approved
The heated persulfate oxidation technique has been approved and adopted in numerous regulatory compliance methods, standards, and norms for water quality testing, including: Standard Method 5310C, USEPA 415.3, USP <643> / EU 2.2.44, ASTM D 4779 and 4839, USEPA-DBPR, USEPA-SPCC, ISO 8245, and EN 1484.
Applications
The Aurora 1030W TOC Analyzer is ideal for the following applications: drinking water, pharmaceutical cleaning validation, municipal wastewater, groundwater / surface water, process water, boiler feed water and condensate, metal plating solutions, and ultrapure water.
Aurora 1030W Features:
- Wide sample range from 10 ppb to 30,000 ppm
- High sample efficiency – Up to 300 samples daily
- Rinse function eliminates residue from the previous sample preventing blanks and background contamination
- Heated persulfate oxidation technique is compliant with numerous regulatory methods, standards, and norms
- Operates with an intuitive color touchscreen display with Windows® CE-based software
- Options available to process more samples and test various sample types
Supported Test Methods
Method | Sample / Application |
Standard Method 5310C | Drinking Water + Wastewater |
USEPA 415.3 | Drinking Water |
USP <643> / EU 2.2.44 | Purified Water |
ASTM D 4779 | Ultrapure Water |
ASTM D 4839 | Wastewater + Seawater |
USEPA-DBPR | Disinfection Byproduct Rule |
USEPA-SPCC | Spill Prevention & Control Countermeasures |
ISO 8245 | Drinking Water & Wastewater |
EN 1484 | Surface & Ground Waters, Potable Water |
Specificaties
Operating Principle | Heated sodium persulfate oxidation |
Measurement Range | 10 ppb C – 30,000 ppm C (multiple calibration ranges or dilution required) |
Instrument Detection LImit (IDL) | 2 ppb C |
Sample Injection Volume | 10µL – 10mL |
Method TC | Acid and persulfate reaction |
Method TIC | Acidification with Phosphoric acid and sparging |
Method TOC | NPOC by heated persulfate oxidation or TC – TIC |
Reagents Required | Sodium persulfate, 5% phosphoric acid, rinsewater |
Gas Supply | N2 (99.998%), zero-grade air, or O2 (99.998%) |
Measurement Technique | Non-dispersive infrared (NDIR) detector |
Operator Interface | Color LCD touchscreen display with Windows® CE-based software |
Basic Software | Single instrument operation with data transfer to PC |
Optional ATOC Software | Network LAN/LIMS operation, data management, custom reports, 21 CFR 11 compliance |
Autosampler | 88 position rotary autosampler designed to fit directly underneath Aurora 1030D analyzer |
Certification | CE, EMC: EN61326 / Safety: IEC 61010-11 2001 |
Power Supply | Variable voltage, 100-240VAC, 50/60Hz, 950W |
Dimensions – Aurora 1030D + 1088 Autosampler | 26.75 in. H x 19.5 in. W x 23 in. D. |