ALOHA+ H2O
The superior analytical solution for your High-Brightness LED needs
The ALOHA+ H2O advances moisture detection in ammonia (NH3) to unprecedented levels. The analyzer provides specialty gas and High-Brightness LED makers with the exceptional detection limits, accuracy, reliability, speed of response and ease of operation that Tiger Optics customers have come to expect. LED manufacturers rely on Tiger Optics’ family of proven Continuous-Wave Cavity Ring-Down Spectroscopybased moisture sensors to ensure the ammonia process gas is of the high quality necessary to produce High Brightness LEDs.
The cost effective analyzer is quick to install, easy to use and effortless to maintain. There are no offline periodic sensor maintenance procedures, no span calibrations, no purifier replacement and no pump rebuilds required. The ALOHA+ H2O analyzer is fully self-calibrating and the “bright” choice for your detection needs!
Description
Designed for trace level moisture analysis in ammonia, the ALOHA+ H2O offers:
- 3 parts per billion (ppb) detection capability in ammonia (NH3) and sub-ppb in inert matrices
- Rapid response time
- Wide dynamic range
- Absolute and calibration-free measurement technology
- Extremely low cost of ownership
- User-friendly touchscreen and remote communication
Applicaties
Detectie
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
H2O in NH3 | 0 – 20 ppm | 3 ppb | 1 ppb |
H2O in N2 | 0 – 6 ppm | 0.5 ppb | 0.2 ppb |
H2O in He | 0 – 3 ppm | 0.3 ppb | 0.1 ppb |
H2O in Ar | 0 – 4 ppm | 0.4 ppb | 0.15 ppb |
Specificaties
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/8h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 5 minutes to 95% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions* | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Gas connections | 1/4″ male VCR inlet and outlet |
Leak tested to | 1 x 10-9 mbar l / sec |
Inlet pressure | 10 – 125 psig (1.7 – 9.6 bara) |
Outlet pressure | < 2 Torr (2.7 mbar) |
Flow rate | Up to 1.8 slpm (gas dependent) |
Sample gases | Ammonia (NH3) and inert matrices |
Gas temperature | Up to 60°C |
Dimensions, H x W x D | |
Standard sensor (incl. shutoff valves) | 8.73″ x 8.57″ x 26.4″ (222 mm x 218 mm x 670 mm) |
Sensor rack (fits up to two sensors) | 8.73″ x 19.0″ x 26.4″ (222 mm x 483 mm x 670 mm) |
Weight | |
Standard sensor | 34 lbs (15.4 kg) |
Electrical and Interfaces | |
Platform | Max series analyzer |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 – 240 VAC, 50/60 Hz |
Power consumption | 40 Watts max. |
Signal output | Isolated 4–20 mA per sensor |
User interfaces | 5.7″ LCD touchscreen, 10/100 Base-T Ethernet |
USB, RS-232, RS-485, Modbus TCP (optional) | |
Data storage | Internal or external flash drive |
Certification | CE Mark |
Patents | |
U.S. Patent #7,277,177 | |
*Vacuum source required |