Portable Gas Sample Probe APP

Due to various acid dew points at various applications the probe is heated up to 180°C.

A filter element of 150mm is integrated, suitable for most applications up to 1g dust/m3. A significant advantage is that this filter is replaceable without dismounting the probe.
With the filter element of 150mm length this probe is suitable for most applications up to 1g dust/m3; with the blow-back function dust loads of up to 10g/m3 can be handled.

Calibration gas can be injected directly into the probe through the calibration gas connection according to EN14181 (regulation for calibration of emission monitoring systems) that enables calibration gas feeding via the filter element of the gas sample probe. The use of the optional high-flow check valve type APP 200 is recommended.

The portable gas sample probe series APP is a perfect completion to the Ankersmid portable gas conditioning system series APS 303/313. Both devices can easily be connected together by the Ankersmid heated sample line type AHL205/025 with the incorporated innovative Quick-Lock System which guarantees a fast and reliable connection.

Using this heated line the probe’s electrical power is then fed to the probe via the feed line integrated within the sample line.
To avoid cooling-down and condensation of sample gas in the extraction area, the heated sample probe tube type AST 301 is available.

Description

Application

The APP portable gas sample probe is especially designed for portable applications where easy handling, safe and reliable operation as well as easy maintenance at a wide range of conditions is required. The low weight and innovative design guarantees a perfect operation for mobile measurements.

Brochure

Ask a quote