nCLD 844 Mhr
ECO PHYSICS nCLD 844 Mhr
Modular Gas Analyzer
The nCLD 844 Mhr analyzer is the next generation in two-channel high precision nitrogen oxide measurement. Unique in speed and reliability, the nCLD 844 Mhr is modular designed and capable of simultaneously measuring NO, NO2 and NOX from hot and humid gas sources without additional cooler. The new and intuitive graphical user interface also individually displays and connects to other instruments’ data.
Description
Specifications
Analyzer type | dual chamber CLD with cooled PMT for measurement of NO, NO2 and NOX |
Measuring ranges | four freely selectable ranges from 0,5 ppm – 500 ppm |
Min. detectable concentration* | 0.012 ppm |
Noise at zero point (1σ)* | 0.006 ppm |
Lag time | <3 sec |
Rise time (0-90%) | <1 sec |
Temperature range | 0-40 °C |
Humidity tolerance | 5-95% rel. h (non-condensing, ambient air and sample gas) |
Dry air use for O3 generator | internally generated (no external supply gas required) |
Sample flow rate | 1.0 l/min |
Input pressure | 600 – 1’200 mbar abs. |
Power required | 350 VA (incl. membrane pump and scrubber) |
Supply voltage | 100-240 V/50-60 Hz |
Interface | USB(3x), HDMI, Bluetooth, RS232 (w/o 9pin connector), LAN, WLAN |
Dimensions | height: 133 mm (5¼”)
width: 450 mm (19”) with molding: 495mm depth: 540 mm (21.2”) |
Weight | 23 kg (51 lb) |
Delivery includes | nCLD 844 Mhr analyzer, power cable, USB-LAN adapter |
Standard | nCLD 844 Mhr: NO, NO2, NOX analyzer
· M – metal converter · h – hot tubing · r – electro-mechanical pressure regulation |
Options | · USB-RS232 9pin connector
· analog output (external box): 0-10 V ; 4-20 mA into 500 Ω max. |
*Depending on filter setting. | |
Connectivity properties are country-specific. | |
ECO PHYSICS reserves the right to change these specifications without notice. |